Safety and efficiency evaluation of an innovative plasma jet array in argon by gas switching technology
Plasma Study
Plasma-kg
conductedBy
- Balazinski, Martina Person
- Gerling, Torsten Person
- Glitsch, Sven Person
- Hahn, Veronika Person
- Horn, Stefan Person
- Lüttjohann, Paul Person
- Timm, Michael Person
- Turski, Philipp Person
- Wagner, Robert Person
- Weltmann, Klaus-Dieter Person
- von Woedtke, Thomas Person
hasMedium
- Ar Medium
- ambient air Medium
hasOutput
hasPlasmaSource
- Plasma Jet Array System (PJAS) Device I Plasma Source
- Plasma Jet Array System (PJAS) Device II Plasma Source
- kINPen-MED Plasma Source
hasTarget
- chronic wounds Target
- conductive metal mesh Target
- copper plate Target
- counter electrode Target
- dielectric plate Target
- dielectric surface Target
- grounded metallic mesh Target
- non-healing wounds Target
- quartz plate Target
hasTopic
- Safety and efficiency evaluation Research Topic
- Wound healing Research Topic
- antibacterial Research Topic
- biomedical applications Application Field
- treating chronic wounds Research Topic
investigates
involves
- chronic wounds Target
- cold atmospheric plasma (CAP) Plasma
- dielectric plate Target
- grounded metallic mesh Target
- non-healing wounds Target
- plasma effluent Plasma
- quartz plate Target
usesConfiguration
- Configuration for PJAS I-1: Device I (homemade integrated circuit, 66 V DC from base unit); Valve Setting 1 (8 ms flushing per channel, 8.2-8.3 ms programming delay); Frequency ~1 MHz; Voltage 3-6 kV ptp; Pressure 1 atm; Argon pre-pressure 5 bar; Argon flow rate 2.5 slm or 5 slm
- Configuration for PJAS I-2: Device I (homemade integrated circuit, 66 V DC from base unit); Valve Setting 2 (30 ms valve opening, 2 ms delay); Frequency ~1 MHz; Voltage 3-6 kV ptp; Pressure 1 atm; Argon pre-pressure 5 bar; Argon flow rate 2.5 slm or 5 slm
- Configuration for PJAS II-2: Device II (homemade discrete circuitry, 60 V DC from base unit); Valve Setting 2 (30 ms valve opening, 2 ms delay); Frequency ~1 MHz; Voltage 3-6 kV ptp; Pressure 1 atm; Argon pre-pressure 5 bar; Argon flow rate 2.5 slm or 5 slm
- Configuration for kINPen® MED: Frequency ~1 MHz; Voltage 3-6 kV ptp; Pressure 1 atm; Argon pre-pressure 5 bar; Argon flow rate 2.5 slm or 5 slm
usesDevice
- APNA-370, Horiba Diagnostic Device
- APOA-360, Horiba Diagnostic Device
- AvaSpec 3648, Avantes Diagnostic Device
- F-201 CM with E-8412-A-10, Bronkhorst Diagnostic Device
- MTS50/M-Z8 with TSC 001 motor control, Thorlabs Diagnostic Device
- TS5, Weidmann Diagnostic Device
- UNIMET 800ST, Bender Diagnostic Device
- base unit Device
- linear stages Device
- motorized stages Device
usesMaterialEntity
- dielectric plate Target
- dielectric surface Target
- grounded metallic mesh Target
- quartz plate Target
usesMethod
- Fiber optic temperature (FOT) analysis Diagnostic Method
- Flow switching technology Method
- Gas switching technology Method
- Optical Emission Spectroscopy (OES) Diagnostic Method
- Ozone and nitrogen dioxide density measurement Diagnostic Method
- Patient leakage current (PLC) measurement Diagnostic Method
- UV absorption spectroscopy Diagnostic Method